202 Agilent X-Series Signal Generators SCPI Command Reference
Analog Modulation Commands
Amplitude Modulation Subsystem–Option UNT ([:SOURce])
This command sets the couplings between start- to- stop values and stop-to- start values for AM sweep
times and rates. The sweep times coupled to on.
Use this command when :AM[1]|2:INTernal:SWEep:SHAPe is set to TRIangle.
ON|1 This choice uses the same sweep time for both start freq to stop freq and stop
freq to start freq sweeps of a triangle shape sweep. If sweep times coupled is ON,
sweep time and rate is only set by the Start -> Stop softkeys.
OFF|0 This choice disables sweep time couplings.
*RST OFF
:AM[1]|2:INTernal:SWEep:TIME[1]|2
Supported All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:TIME[1]|2 <val><unit>
[:SOURce]:AM[1]|2:INTernal:SWEep:TIME[1]|2?
This command sets the sweep rate for the amplitude- modulated, swept- sine waveform.
*RST
+1.00000000Eï€001
Range 1mS–65.535S
Key Entry AM Sweep Time
:AM[1]|2:INTernal:SWEep:TRIGger
Supported All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:TRIGger BUS|IMMediate|EXTernal|INTernal|KEY
[:SOURce]:AM[1]|2:INTernal:SWEep:TRIGger?
This command sets the trigger source for the amplitude modulated swept- sine waveform.
BUS This choice enables GPIB triggering using the *TRG or GET command or LAN and
RS- 232 triggering using the *TRG command.
IMMediate
This choice enables immediate triggering of the sweep event.
EXTernal This choice enables the triggering of a sweep event by an externally applied signal at the
TRIG 1, TRIG 2, or PULSE BNC connector.
INTernal This choice enables the triggering of a sweep event by an internal Pulse Video or Pulse
Sync signal.
KEY This choice enables triggering through front panel interaction by pressing the
Trigger hardkey.
*RST
IMM
Key Entry AM Sweep Times Coupled Off On
Key Entry Bus Free Run Ext Int Trigger Key