Agilent X-Series Signal Generators SCPI Command Reference 143
System Commands
Output Subsystem (:OUTPut)
Output Subsystem (:OUTPut)
:BLANking:AUTO
Supported All Models
:OUTPut:BLANking:AUTO ON|OFF|1|0
:OUTPut:BLANKing:AUTO?
This command turns the RF output on or off during frequency band changes. Frequency band
changes can cause the signal generator’s RF output to fluctuate. The output blanking function, when
active, turns off the RF output until the frequency and power settles.
ON(1) The RF output turns off when crossing a frequency band.
OFF(0) The RF output stays on, if possible, when crossing a frequency band. Refer to the
Data sheet.
*RST 1
Key Entry Output Blanking Off On Auto
Remarks Refer to the signal generator’s data sheet for information on frequency switching
speeds, settling times, and frequency band information.
:BLANking:STATe
Supported All Models
:OUTPut:BLANking:STATe ON|OFF|1|0
:OUTPut:BLANKing:STATe?
This command enables or disables the RF output blanking state.
ON(1) The RF output turns off during frequency changes.
OFF(0) The RF output stays on, if possible, during frequency changes. Refer to the Data
sheet.
*RST 0
Remarks Refer to the signal generator’s data sheet for information on frequency switching
speeds, settling times, and frequency band information.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals