Agilent X-Series Signal Generators SCPI Command Reference 23
Basic Function Commands
Correction Subsystem ([:SOURce]:CORRection)
Remarks The setting enabled by this command is not affected by signal generator power–on,
preset, or *RST.
:PMETer:COMMunicate:LAN:DEVice
Supported All Models
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:DEVice <deviceName>
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:DEVice?
This command enters a VXI–11 device name for a power meter that is being controlled by the signal
generator for user flatness calibration. If connecting directly to the power meter, enter the name as
specified on your power meter documentation. If connecting through a LAN–GPIB gateway, enter the
SICL address of the power meter.
Key Entry PM VXI–11 Device Name
Remarks The setting enabled by this command is not affected by signal generator power–on,
preset, or *RST.
:PMETer:COMMunicate:LAN:IP
Supported All Models
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:IP <ipAddress>
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:IP?
This command sets the internet protocol (IP) address for a power meter that is controlled by the
signal generator for user flatness calibration. If connecting to a GPIB power meter through a
LAN–GPIB gateway, this command sets the IP address of the gateway.
Key Entry Power Meter IP Address
Remarks The setting enabled by this command is not affected by signal generator power–on,
preset, or *RST.
Ensure that the power meter IP address is different from the signal generator
address.
:PMETer:COMMunicate:LAN:PORT
Supported All Models
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:PORT <portNumber>
[:SOURce]:CORRection:PMETer:COMMunicate:LAN:PORT?
This command sets the IP port number on the power meter that is controlled by the signal generator
for users flatness calibration.
Key Entry Power Meter IP Port
5025 Standard mode. The command enables standard mode for simple programming.
5024 Telnet mode. The command enables the telnet SCPI service for programming.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals