Agilent X-Series Signal Generators SCPI Command Reference 327
Arb Commands
Multitone Subsystem–N5172B/82B ([:SOURce]:RADio:MTONe:ARB)
:RAD:MTON:ARB:SET:TABL:ROW 2,-10,40,0
The preceding example modifies row number two in the currently selected multitone table. The power
is set to –10 dB, the phase is set to 40 degrees, and the state is off.
[:STATe]
Supported N5172B/82B
[:SOURce]:RADio:MTONe:ARB[:STATe] ON|OFF|1|0
[:SOURce]:RADio:MTONe:ARB[:STATe]?
This command enables or disables the multitone waveform generator function.
*RST 0
Key Entry Multitone Off On
*RST frequency offset: 3.50000000E+004 <power>: +0.00000000E+000
<phase>: +0.00000000E+000 <state>: 1
Range frequency offset::
653: +/- 30 MHz
655: +/- 60 MHz
656: +/- 40 MHz
657: +/- 80 MHz
<power>: 80 to
0
<phase>: 0–359 <state>: 1
Key Entry Goto Row Edit Item Toggle State