Agilent X-Series Signal Generators SCPI Command Reference 77
Basic Function Commands
Power Subsystem ([:SOURce]:POWer)
Key Entry Amptd Offset
Remarks This simulates a power level at a test point beyond the RF OUTPUT connector
without changing the actual RF output power. The offset value only affects the
displayed amplitude setting.
You can enter an amplitude offset any time in either normal operation or
amplitude reference mode.
[:LEVel][:IMMediate][:AMPLitude]
Supported All Models
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude] <value><unit>
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]?
This command sets the RF output power.
Key Entry AMPTD
Remarks For information on the ranges for this command and the specified values, refer to
the instrument’s Data Sheet.
[:LEVel]:MINimum:LIMit
Supported All Models w/Option HAL
[:SOURce]:POWer[:LEVel]:MINimum:LIMit LOW|HIGH
[:SOURce]:POWer[:LEVel]:MINimum:LIMit?
*RST –1.10000000E002 (Standard) or 1.44000000E002 (Option 1EQ)