Agilent X-Series Signal Generators SCPI Command Reference 165
System Commands
System Subsystem (:SYSTem)
:ERRor[:NEXt]
Supported All Models
:SYSTem:ERRor[:NEXt]?
This query returns the next error message from the signal generator SCPI error queue. If there are
no error messages, the query returns the following output:
+0,"No error"
When there is more than one error message, the query will need to be sent for each message.
The Agilent MXG deletes the error messages from the front- panel error queue after viewing the last
message.
:ERRor:SCPI[:SYNTax]
Supported All Models
:SYSTem:ERRor:SCPI[:SYNTax] ON|OFF|1|0
:SYSTem:ERRor:SCPI[:SYNTax]?
This command enables or disables the reporting of SCPI syntax errors to the error queue.
The setting ON/1 is persistent through Preset and *RST. It is cleared to OFF/0 by sending
SYST:PRES:ALL or cycling the power of the signal generator.
:FILesystem:STORage:EXTernal
Supported All Models
:SYSTem:FILesystem:STORage:EXTernal?
This query checks to see if the external USB port is actively being used for data storage and retrieval
on the signal generator. A returned value of 1 means the external USB media is being used for data
storage and retrieval. For more information on non–volatile storage media settings, refer to
:FILesystem:STORage:EXTernal:PATH, :FILesystem:STORage:TYPE and
:FILesystem:STORage:TYPE:AUTO commands.
:FILesystem:STORage:EXTernal:PATH
Supported All Models
:SYSTem:FILesystem:STORage:EXTernal:PATH <"USB media root path">
:SYSTem:FILesystem:STORage:EXTernal:PATH?
This command selects the directory storage path on the USB media. For more information, refer to
the signal generator’s softkey Help. For more information on non–volatile storage media settings, refer
to :FILesystem:STORage:EXTernal, :FILesystem:STORage:TYPE and :FILesystem:STORage:TYPE:AUTO
commands.
Key Entry Error Info View Next Error Message