Agilent X-Series Signal Generators SCPI Command Reference 175
System Commands
System Subsystem (:SYSTem)
:PMETer[1]|2:COMMunicate:LAN:PORT
Supported All Models
:SYSTem:PMETer[1]|2:COMMunicate:LAN:PORT <portNumber>
:SYSTem:PMETer[1]|2:COMMunicate:LAN:PORT?
This command sets the IP port on the power meter that is controlled by the signal generator.
Key Entry Power Meter IP Port
5025 Standard mode. The command enables standard mode for simple programming.
5024 Telnet mode. The command enables the telnet SCPI service for programming.
NOTE For firmware versions <A.01.51, the default telnet mode is 5023. For firmware versions
A.01.51 and greater, telnet port 5023 is still available for backwards compatibility.
Remarks The setting enabled by this command is not affected by signal generator power–on,
preset, or *RST.
For more information on standard mode and telnet SCPI mode, refer to the
Programming Guide.
:PMETer[1]|2:COMMunicate:TYPE
Supported All Models
:SYSTem:PMETer[1]|2:COMMunicate:TYPE SOCKets|SOCKETS|VXI11|USB
:SYSTem:PMETer[1]|2:COMMunicate:TYPE?
This command sets the type of control connection for communication with the external power meter
for power meter measurements. The query returns the connection type.
Key Entry Connection Type
SOCK or
SOCKETS The command enables the power meter for sockets LAN control through the signal
generator.
VXI11 The command enables the power meter for VXI–11 control through the signal
generator. A power meter with GPIB can be controlled through VXI–11 using a
LANGPIB gateway.
USB The command enables the power meter for USB control through the signal
generator.
Remarks The setting enabled by this command is not affected by signal generator power–on,
preset, or *RST.