122 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Memory Subsystem (:MEMory)
The signal generator will return the two memory usage parameters and as many file listings as there
are files in the memory subsystem. Each file listing parameter will be in the following form:
"<file name,file type,file size>"
Example Output
1818624,519962624,0_00@STATE,STAT,641","0_01@STATE,STAT,669","A@NVHDR,NVHDR,132","A@N
VMKR,NVMKR,0","A@NVWFM,NVWFM,9","COPY12@STATE,STAT,669","FLAT_1@USERFLAT,UFLT,16","GE
N_FILE11@BINARY,BIN,5","LAST@LIST,LIST,122","LAST@USERFLAT,UFLT,16","PERSISTENT@STATE
,STAT,1056",SEQ1_TEST@SEQ,SEQ,206
Key Entry All
Remarks Refer to the Table 1- 4 on page 15 for a listing of the file types and the table on
page 15 for information on the "<file name>" syntax.
:COPY[:NAME]
Supported All Models
:MEMory:COPY[:NAME] "<file name>","<file name>"
This command makes a duplicate of the requested file.
Key Entry Copy File
Remarks Refer to File Name Variables for information on the file name syntax.
When copying a waveform file from volatile to non–volatile memory, the marker
file and file header, associated with the waveform file, will automatically be copied
at the same time.
:DATA
Supported All Models
:MEMory:DATA "<file_name>",<data_block>
:MEMory:DATA? "<file_name>"
This command loads data into signal generator memory using the <data_block> parameter and saves
the data to a file designated by the "<file_name>" variable. The query returns the file contents of the
file as a datablock.
A waveform file must be located in volatile waveform memory (WFM1) before it can be played by the
signal generator’s dual ARB player.
For downloads directly into volatile waveform memory (WFM1) use the path "WFM1:<file_name>". For
downloads to non–volatile waveform memory, use the path "NVWFM:<file_name>".
"<file_name>" This variable names the destination file, including the directory path.
<data_block> This parameter represents the data and file length parameters. The data in the file
is represented by the <data_block> variable.
Refer to the Programming Guide for more information on programming the status registers.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals