116 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
IEEE 488.2 Common Commands
*STB?
Supported All Models
*STB?
The Read Status Byte (STB) query returns the value of the status byte including the master summary
status (MSS) bit.
Range 0 to 255
Remarks Refer to the Programming Guide for more information.
*TRG
Supported All Models
*TRG
The Trigger (TRG) command triggers the device if BUS is the selected trigger source, otherwise, *TRG
is ignored.
*TST?
Supported All Models
*TST?
The Self–Test (TST) query initiates the internal self–test and returns one of the following results:
0 This shows that all tests passed.
1 This shows that one or more tests failed.
Key Entry Run Complete Self Test
*WAI
Supported All Models
*WAI
The Wait–to–Continue (WAI) command causes the signal generator to wait until all pending
commands are completed, before executing any other commands.