Agilent X-Series Signal Generators SCPI Command Reference 137
System Commands
Memory Subsystem (:MEMory)
:LOAD:LIST
Supported All Models
:MEMory:LOAD:LIST "<file name>"
This command loads a list sweep file.
Key Entry Load From Selected File
:LOAD:PTRain
Supported All with Option 320 and either UNU or UNW
:MEMory:LOAD:PTRain <"filename">
This command reads the pulse train file specified. Refer to :STORe:PTRain.
Key Entry Confirm Load from File
:MOVE
Supported All Models
:MEMory:MOVE "<src_file>","<dest_file>"
This command renames the requested file in the memory catalog.
Key Entry Rename File
Remarks Refer to File Name Variables for information on the file name syntax.
:SIZE
Supported All Models
:MEMory:SIZE? <"filename">
This command returns the size of the file named <"filename"> in bytes or a –1, if the file does not
exist. If the MSUS or directory is invalid, an “ERROR: –257, File name error” will be reported.
:STATe:COMMent
Supported All Models
:MEMory:STATe:COMMent <reg_num>,<seq_num>,"<comment>"
:MEMory:STATe:COMMent? <reg_num>,<seq_num>
This command lets you to add a descriptive comment to the saved state <reg_num>,<seq_num>.
Comments can be up to 55 characters long.
Key Entry Add Comment To Seq[n] Reg[nn]
:STORe:LIST
Supported All Models