EasyManua.ls Logo

Agilent Technologies N5171B

Agilent Technologies N5171B
458 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent X-Series Signal Generators SCPI Command Reference 137
System Commands
Memory Subsystem (:MEMory)
:LOAD:LIST
Supported All Models
:MEMory:LOAD:LIST "<file name>"
This command loads a list sweep file.
Key Entry Load From Selected File
:LOAD:PTRain
Supported All with Option 320 and either UNU or UNW
:MEMory:LOAD:PTRain <"filename">
This command reads the pulse train file specified. Refer to :STORe:PTRain.
Key Entry Confirm Load from File
:MOVE
Supported All Models
:MEMory:MOVE "<src_file>","<dest_file>"
This command renames the requested file in the memory catalog.
Key Entry Rename File
Remarks Refer to File Name Variables for information on the file name syntax.
:SIZE
Supported All Models
:MEMory:SIZE? <"filename">
This command returns the size of the file named <"filename"> in bytes or a1, if the file does not
exist. If the MSUS or directory is invalid, anERROR: –257, File name error” will be reported.
:STATe:COMMent
Supported All Models
:MEMory:STATe:COMMent <reg_num>,<seq_num>,"<comment>"
:MEMory:STATe:COMMent? <reg_num>,<seq_num>
This command lets you to add a descriptive comment to the saved state <reg_num>,<seq_num>.
Comments can be up to 55 characters long.
Key Entry Add Comment To Seq[n] Reg[nn]
:STORe:LIST
Supported All Models

Table of Contents