20 Agilent X-Series Signal Generators SCPI Command Reference
Basic Function Commands
Correction Subsystem ([:SOURce]:CORRection)
Correction Subsystem ([:SOURce]:CORRection)
:FLATness:FREQuency
Supported All Models
[:SOURce]:CORRection:FLATness:FREQuency? <point>
This command returns the frequency value of the <point> queried.
Range 1 to 3,201
Key Entry # Points
:FLATness:INITialize:FSTep
Supported All Models
CAUTION The current flatness data will be overwritten once this command is executed. If needed,
save the current data. Refer to the :FLATness:STORe command for storing user flatness
files.
[:SOURce]:CORRection:FLATness:INITialize:FSTep
This command replaces the loaded user flatness data with the settings from the current step array
data points.
The maximum number of user flatness points is 3,201. When copying the step array settings over to
a user flatness file, ensure that the number of points in the step array do not exceed the maximum
user flatness points.
See also “:FLATness:STEP:POINts” on page 21.
Range 1 to 3,201
Key Entry Load Cal Array From Step Array
:FLATness:LOAD
Supported All Models
[:SOURce]:CORRection:FLATness:LOAD "<file name>"
This command loads a user–flatness correction file. The "<file name>" variable is the name of the file
located in the Catalog of USERFLAT Files. The directory path is implied in the command and need
not be specified in the file name parameter. For more information on file name syntax, refer to “File
Name Variables” on page 13.
Key Entry Load From Selected File