Agilent X-Series Signal Generators SCPI Command Reference 47
Basic Function Commands
Frequency Subsystem ([:SOURce])
:PHASe[:ADJust]
Supported All Models
[:SOURce]:PHASe[:ADJust] <value><unit>
[:SOURce]:PHASe[:ADJust]?
This command adjusts the phase of the modulating signal.
The query will only return values in radians.
*RST 0.00000000E000
Key Entry Adjust Phase
:ROSCillator:BANDwidth:EXTernal
Supported All Models
[:SOURce]:ROSCillator:BANDwidth:EXTernal
<value>[<units>]|NARRow|WIDE|MINimum|MAXimum|DEFault
[:SOURce]:ROSCillator:BANDwidth:EXTernal? |MINimum|MAXimum|
This command selects the external frequency bandwidth as the source for the measurement.
For values greater than 9.5 Hz, 73 Hz is used.
*RST +9.50000000E+000
Range .5 or 73 Hz
Key Entry Ref Oscillator Ext Bandwidth
:ROSCillator:FREQuency:BBG
Supported All Models
[:SOURce]:ROSCillator:FREQuency:EXTernal <value>
[:SOURce]:ROSCillator:FREQuency:EXTernal?
This command sets the frequency of the internal baseband generator reference oscillator.
*RST +1.0000000000000E+07 Hz
Range +1.0000000000000E+06 to +5.0000000000000E+07 Hz
Key Entry Ref Oscillator Ext Freq
:ROSCillator:FREQuency:EXTernal
Supported All Models
[:SOURce]:ROSCillator:FREQuency:EXTernal <value>
[:SOURce]:ROSCillator:FREQuency:EXTernal?
This command makes External Ref Frequency the active function. The value that you enter sets the
frequency of the external reference oscillator.
Range Radians: –3.14 to 3.14 RAD Degrees: –180 to 179 DEG