Agilent X-Series Signal Generators SCPI Command Reference 229
Analog Modulation Commands
Phase Modulation Subsystem–Option UNT ([:SOURce])
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:SHAPe?
This command sets the shape of tone 1 (default) or tone 2 of the internal dual function generator
source.
*RST SINE
:PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:SHAPe:RAMP
Supported All Models with Option UNT and 303
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:SHAPe:RAMP POSitive|NEGative
[:SOURce]:PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:SHAPe:RAMP?
This command sets the ramp direction of the selected tone (1 or 2) of the internal dual function
generator source when :PM[1]|2:INTernal:DUAL:FUNCtion[1]|2:SHAPe is set to RAMP.
*RST POS
:PM:INTernal:FREQuency:STEP[:INCRement]
Supported All Models with Option UNT
[:SOURce]:PM:INTernal:FREQuency:STEP[:INCRement] <num>
[:SOURce]:PM:INTernal:FREQuency:STEP[:INCRement]?
This command sets the step increment of the phase modulation internal frequency.
The variable <num> sets the entered value in units of Hertz.
*RST +5.00000000E+002 (persistent value; use :PRESet:PERSistent to restore the factory
preset value)
Range 0.1–6.25E6
Key Entry Incr Set
Remarks The value set by this command is used with the UP and DOWN choices for the PM
frequency command. Refer to :PM[1]|2:INTernal:FUNCtion[1]|2:FREQuency
command for more information.
The setting enabled by this command is not affected by signal generator power- on,
preset, or *RST.
:PM[1]|2:INTernal:FUNCtion[1]|2:FREQuency
Supported All Models with Option UNT
[:SOURce]:PM[1]|2:INTernal:FUNCtion[1]|2:FREQuency <value><unit>|UP|DOWN
[:SOURce]:PM[1]|2:INTernal:FUNCtion[1]|2:FREQuency?
This command sets the internal modulation frequency rate for the following applications:
• the start frequency for a swept- sine waveform
Key Entry
M Tone 1 Waveform M Tone 2 Waveform
Key Entry Pos Ramp Neg Ramp