402 Agilent X-Series Signal Generators SCPI Command Reference
Real–Time Commands
Custom Subsystem–Option 431([:SOURce]:RADio:CUSTom)
:STANdard:SELect
Supported N5172B/82B with Option 431
[:SOURce]:RADio:CUSTom:STANdard:SELect NONE|AC4Fm|ACQPsk|BLUEtooth|CDPD
[:SOURce]:RADio:CUSTom:STANdard:SELect?
This command selects a predefined setup for Custom (with the appropriate defaults) and/or clears
the selection.
NONE
This choice clears the current predefined Custom format.
AC4Fm This choice sets up an Association of Public Safety Communications Officials (APCO)
compliant, compatible 4-level frequency modulation (C4FM) format.
ACQPsk This choice sets up an Association of Public Safety Communications Officials (APCO)
compliant, compatible quadrature phase shift keying (CQPSK) format.
BLUEtooth This choice sets up a Bluetooth (2-level frequency shift keying) format.
CDPD This choice sets up a minimum shift keying Cellular Digital Packet Data (CDPD)
format.
*RST NONE
:TRIGger:TYPE
Supported N5172B/82B with Option 431
[:SOURce]:RADio:CUSTom:TRIGger:TYPE CONTinuous|SINGle|GATE
[:SOURce]:RADio:CUSTom:TRIGger:TYPE?
This command sets the trigger type.
CONTinuous
The framed data sequence repeats continuously; the sequence restarts every time the
previous playback is completed. To customize continuous triggering, refer to
“:TRIGger:TYPE:CONTinuous[:TYPE]” on page 402.
SINGle The framed data sequence plays once for every trigger received.
GATE An external trigger signal interrupts the playback while the gating signal is in the
inactive state. Playback resumes when the external control signal returns to the active
state. The active state can be set to high or low.
*RST CONT
:TRIGger:TYPE:CONTinuous[:TYPE]
Supported N5172B/82B with Option 431
[:SOURce]:RADio:CUSTom:TRIGger:TYPE:CONTinuous[:TYPE] FREE|TRIGger|RESet
[:SOURce]:RADio:CUSTom:TRIGger:TYPE:CONTinuous[:TYPE]?
Key Entry None APCO 25w/C4FM APCO 25 w/CQPSK Bluetooth CDPD
Key Entry Continuous Single Gated

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals