52 Agilent X-Series Signal Generators SCPI Command Reference
Basic Function Commands
List/Sweep Subsystem ([:SOURce])
:LIST:DWELl:TYPE
Supported All Models
[:SOURce]:LIST:DWEL1:TYPE LIST|STEP
[:SOURce]:LIST:DWEL1:TYPE?
This command toggles the dwell time for the list sweep points between the values defined in the list
sweep and the value for the step sweep.
LIST This choice selects the dwell times from the list sweep. Refer to the :LIST:DWELl
command for setting the list dwell points.
STEP This choice selects the dwell time from the step sweep. Refer to the
:SWEep:DWELl command for setting the step dwell.
*RST LIST
Key Entry Dwell Type List Step
:LIST:FREQuency
Supported All Models
[:SOURce]:LIST:FREQuency <value>,<value>
[:SOURce]:LIST:FREQuency?
This command sets the frequency values for the current list sweep points.
The maximum number of list sweep points is 3,201.
The variable <value> is expressed in units of hertz (Hz).
The setting enabled by this command is not affected by signal generator power–on, preset, or *RST.
Range The range is model/option dependent. Refer to the instrument’s Data Sheet.
:LIST:FREQuency:POINts
Supported All Models
[:SOURce]:LIST:FREQuency:POINts?
This command queries the current list sweep file for the number of frequency points.
:LIST:MANual
Supported All Models
[:SOURce]:LIST:MANual <value>|UP|DOWN
[:SOURce]:LIST:MANual?
This command sets a list or step sweep point as the current sweep point controlling the frequency
and power output.
If list or step mode is controlling frequency or power, or both, then the indexed point in the
respective list(s) will be used.