22 Agilent X-Series Signal Generators SCPI Command Reference
Basic Function Commands
Correction Subsystem ([:SOURce]:CORRection)
:FLATness:STEP:STARt
Supported All Models
[:SOURce]:CORRection:FLATness:STEP:STARt <freq><unit> |MAXimum|MINimum|DEFault|
[:SOURce]:CORRection:FLATness:STEP:STARt? [MAXimum|MINimum]
This command sets the start frequency for the user flatness calibration step array. See also,
:FLATness:STEP:POINts and :FLATness:STEP:STOP commands.
*RST The preset value is model/option dependent. Refer to the instrument’s Data Sheet.
Range The range is model/option dependent. Refer to the instrument’s Data Sheet.
Key Entry Freq Start
:FLATness:STEP:STOP
Supported All Models
[:SOURce]:CORRection:FLATness:STEP:STOP <freq><unit> |MAXimum|MINimum|DEFault|
[:SOURce]:CORRection:FLATness:STEP:STOP? [MAXimum|MINimum]
This command sets the stop frequency for the user flatness calibration step array. See also,
:FLATness:STEP:POINts and :FLATness:STEP:STARt commands.
*RST The preset value is model/option dependent. Refer to the instrument’s Data Sheet.
Range The range is model/option dependent. Refer to the instrument’s Data Sheet.
Key Entry Freq Stop
:FLATness:STORe
Supported All Models
[:SOURce]:CORRection:FLATness:STORe "<file name>"
This command stores the current user–flatness correction data to a file named by the
:CORRection:FLATness:STORe command. The directory path is implied in the command and need
not be specified in the "<file name>" variable.
Key Entry Store To File
Remarks For information on file name syntax, refer to “File Name Variables” on page 13.
:PMETer:CHANnel
Supported All Models
[:SOURce]:CORRection:PMETer:CHANnel A|B
[:SOURce]:CORRection:PMETer:CHANnel?
This command selects the channel setting on the external power meter for user flatness calibration.
Default Channel A
Key Entry Power Meter Channel A B