110 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Display Subsystem (:DISPlay)
Remarks The display capture is stored as DISPLAY.BMP in the Binary Directory file system.
This file is overwritten with each subsequent display capture. The file can be
down–loaded in the following manner:
1. Log on to the signal generator using ftp.
2. Change (cd) to the BIN directory.
3. Retrieve the file by using the GET command or by using the :MEM:DATA query
on page 122.
:CMAP:DEFault
Supported All Models
:DISPlay:CMAP:DEFault [<palette:BRIGht|DARK|MONOchrome>]
This command selects the color palette for the instrument display.
Remarks The setting enabled by this command is not affected by signal generator power–on,
preset, or *RST.
:REMote
Supported All Models
:DISPlay:REMote ON|OFF|1|0
:DISPlay:REMote?
This command enables or disables the display updating when the signal generator is remotely
controlled.
ON (1) This choice updates the signal generator display (Text Area) so you can see the
settings as the commands are executed, however, this will degrade the signal
generator speed. Frequency Area, Amplitude Area, and status LEDs continue to
update. For more information on the front- panel display description, refer to the
User’s Guide.
OFF (0) This choice turns off the display (Text Area) updating while further optimizing the
signal generator for speed. No Text Area updates occur but the Frequency Area,
Amplitude Area, and status LEDs continue to update. For more information on the
front- panel display description, refer to the User’s Guide.
Key Entry Update in Remote Off On
Remarks The setting enabled by this command is not affected by signal generator preset or
*RST. However, cycling the signal generator power will reset it to zero.
Key Entry Bright Color Dark Color Monochrome