112 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
IEEE 488.2 Common Commands
IEEE 488.2 Common Commands
*CLS
Supported All Models
*CLS
The Clear Status (CLS) command clears the status byte by emptying the error queue and clearing all
the event registers including the Data Questionable Event Register, the Standard Event Status
Register, the Standard Operation Status Register and any other registers that are summarized in the
status byte.
*ESE
Supported All Models
*ESE <data>
The Standard Event Status Enable (ESE) command sets the Standard Event Status Enable Register.
The variable <data> represents the sum of the bits that will be enabled.
Range 0 to 255
Remarks The setting enabled by this command is not affected by signal generator preset or
*RST. However, cycling the signal generator power will reset this register to zero.
Refer to the Programming Guide for more information.
*ESE?
Supported All Models
*ESE?
The Standard Event Status Enable (ESE) query returns the value of the Standard Event Status
Enable Register.
Remarks Refer to the Programming Guide for more information.
*ESR?
Supported All Models
CAUTION This is a destructive read. The data in the register is latched until it is queried. Once
queried, the data is cleared.
*ESR?
The Standard Event Status Register (ESR) query returns the value of the Standard Event Status
Register.
Remarks Refer to the Programming Guide for more information.