218 Agilent X-Series Signal Generators SCPI Command Reference
Analog Modulation Commands
Low Frequency Output Subsystem ([:SOURce]:LFOutput)
[:SOURce]:LFOutput:DUAL:FUNCtion[1]|2:FREQuency?
This command sets the frequency of tone 1 (default) or tone 2 of the internal dual function generator
source.
*RST 400.0 Hz
:DUAL:FUNCtion[1]|2:SHAPe
Supported All Models with Options UNT and 303
[:SOURce]:LFOutput:DUAL:FUNCtion[1]|2:SHAPe SINE|TRIangle|SQUare|RAMP
[:SOURce]:LFOutput:DUAL:FUNCtion[1]|2:SHAPe?
This command sets the shape of tone 1 (default) or tone 2 of the internal dual function generator
source.
*RST SINE
:DUAL:FUNCtion[1]|2:SHAPe:RAMP
Supported All Models with Options UNT and 303
[:SOURce]:LFOutput:DUAL:FUNCtion[1]|2:SHAPe:RAMP POSitive|NEGative
[:SOURce]:LFOutput:DUAL:FUNCtion[1]|2:SHAPe:RAMP?
This command sets the ramp direction of the selected tone (1 or 2) of the internal dual function
generator source when :DUAL:FUNCtion[1]|2:SHAPe is set to RAMP.
*RST POS
:FUNCtion[1]|2:FREQuency
Supported All models with Option UNT
[:SOURce]:LFOutput:FUNCtion[1]|2:FREQuency <val><units>
[:SOURce]:LFOutput:FUNCtion[1]|2:FREQuency?
This command sets the frequency of function generator 1 or 2. The command sets:
the frequency of the first tone of a dual–sine waveform
the start frequency for a swept–sine waveform
the frequency for all other waveform types
For selecting the waveform type, use the :FUNCtion[1]|2:SHAPe command.
*RST +4.00000000E+002
Range 100.0 mHz to 6.25 MHz
Key Entry LF Out Tone 1 Freq LF Out Tone 2 Freq
Key Entry LF Out Tone 1 Waveform LF Out Tone 2 Waveform
Key Entry Pos Ramp Neg Ramp
Range Sine and Dual–Sine: 0.5 Hz to 1 MHz

Table of Contents