Agilent X-Series Signal Generators SCPI Command Reference 153
System Commands
Status Subsystem (:STATus)
Status Subsystem (:STATus)
:OPERation:CONDition?
Supported All Models
:STATus:OPERation:CONDition?
This query returns the decimal sum of the bits for the registers that are set to one and are part of
the Standard Operation Status Group. For example, if a sweep is in progress (bit 3), the value 8 is
returned.
Range 0 to 32767
Remarks The data in this register is continuously updated and reflects current conditions.
Refer to the Programming Guide for more information.
:OPERation:ENABle
Supported All Models
:STATus:OPERation:ENABle <value>
:STATus:OPERation:ENABle?
This command determines which bits in the Standard Operation Event Register will set the Standard
Operation Status Summary bit (bit 7) in the Status Byte Register.
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0 to 32767
Remarks Refer to the Programming Guide for more information.
:OPERation:NTRansition
Supported All Models
:STATus:OPERation:NTRansition <value>
:STATus:OPERation:NTRansition?
This command determines which bits in the Standard Operation Condition Register will set the
corresponding bit in the Standard Operation Event Register when that bit has a negative transition (1
to 0).
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0 to 32767