242 Agilent X-Series Signal Generators SCPI Command Reference
Analog Modulation Commands
Pulse Modulation Subsystem–Option UNU and UNW and Option 320 ([:SOURce])
:PULM:INTernal:TRAin:ONTime
Supported All with Options UNW and 320
[:SOURce]:PULM:INTernal:TRAin:ONTime <value>,<value>
[:SOURce]:PULM:INTernal:TRAin:ONTime?
This command sets the pulse on values for the current list of pulse train on times. If this list is
shorter than the other lists, then the last element will be repeated as necessary to match the length
of the Off Time or the Repetition list. The query returns the count of pulse cycle elements in the list
of on times. Refer to :PULM:INTernal:TRAin:OFFTime and :PULM:INTernal:TRAin:REPetition.
The resolution for this setting is 10 nS.
Example
:PULM:INT:TRA:ONT 100NS,200NS,400E–9
The preceding example sets the pulse train on cycles to 100 nanoseconds, 200 nanoseconds, and 400
nanoseconds.
:PULM:INTernal:TRAin:ONTime:POINts?
Supported All with Options UNW and 320
[:SOURce]:PULM:INTernal:TRAin:ONTime:POINts?
This query returns the count of elements in the list of on times.
:PULM:INTernal:TRAin:REPetition
Supported All with Options UNW and 320
[:SOURce]:PULM:INTernal:TRAin:REPetition <value>,<value>
[:SOURce]:PULM:INTernal:TRAin:REPetition?
This command generates a user- defined list of the pulse repetitions. The maximum is a total count of
2047, so a list of "2047,1" would be too long. Lists that are too long will generate an error and only
the first 2047 pulses will be played. If this list is shorter than the other lists, then the last element
will be repeated as necessary to match the length of the On Time or Off Time list, whichever is
longer. Refer to :PULM:INTernal:TRAin:ONTime and :PULM:INTernal:TRAin:OFFTime.
Example
:PULM:INT:TRA:REP 100,20,3
The preceding example repeats the first pulse cycle 100 times, the second cycle to be repeated 20
times, and the third cycle 3 times.
Range 1 to 2047 total pulse repetitions
Range
120nS to 42S (Option UNU)
20nS to 42S (Option UNW)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals