Agilent X-Series Signal Generators SCPI Command Reference 225
Analog Modulation Commands
Low Frequency Output Subsystem ([:SOURce]:LFOutput)
:SWEep:TRIGger
Supported All Models with Options UNT and 303
[:SOURce]:LFOutput:SWEep:TRIGger IMMediate|KEY|EXTernal|INTernal|BUS
[:SOURce]:LFOutput:SWEep:TRIGger?
This command sets the trigger source for the frequency modulated swept- sine waveform.
IMMediate
This choice enables immediate triggering of the sweep event.
KEY This choice enables triggering through front panel interaction by pressing the
Trigger hardkey.
EXTernal
This choice enables the triggering of a sweep event by an externally applied signal at the
TRIG 1, TRIG 2, or PULSE BNC connector.
INTernal This choice enables the triggering of a sweep event by an internal Pulse Video or Pulse
Sync signal.
BUS This choice enables GPIB triggering using the *TRG or GET command or LAN and
RS- 232 triggering using the *TRG command.
*RST
IMM
:SWEep:TRIGger:EXTernal:SOURce
Supported All Models with Options UNT and 303
[:SOURce]:LFOutput:SWEep:TRIGger:EXTernal:SOURce TRIGger[1]|TRIGger2|PULSe
[:SOURce]:LFOutput:SWEep:TRIGger:EXTernal:SOURce?
This command selects the external trigger source for the sweep. With external triggering, the selected
bi- directional BNC is configured as an input.
TRIGger1 This choice selects the TRIG 1 BNC as the external trigger source for triggering
sweep, point and function generator sweeps.
TRIGger2 This choice selects the TRIG 2 BNC as the external trigger source for triggering
sweep, point and function generator sweeps.
PULSe This choice selects the PULSE BNC as the external trigger source for triggering
sweep, point and function generator sweeps.
*RST TRIGger1
:SWEep:TRIGger:INTernal:SOURce
Supported All Models with Options UNT and 303
[:SOURce]:LFOutput:SWEep:TRIGger:INTernal:SOURce PVIDeo|PSYNc
[:SOURce]:LFOutput:SWEep:TRIGger:INTernal:SOURce?
Key Entry Bus Free Run Ext Int Trigger Key
Key Entry Trigger 1 Trigger 2 Pulse