EasyManuals Logo
Home>Agilent Technologies>Computer Hardware>N5171B

Agilent Technologies N5171B User Manual

Agilent Technologies N5171B
458 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #220 background imageLoading...
Page #220 background image
190 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Trigger Subsystem
:TRIGger:OUTPut[1]|2:POLarity
Supported All
:TRIGger:OUTPut[1]|2:POLarity POSitive|NEGative
:TRIGger:OUTPut[1]|2:POLarity?
Sets the polarity of the TTL signal output at the selected (TRIG 1 or TRIG 2) output BNC.
POSitive TTL high at the start of a dwell sequence, or when waiting for the point trigger in
manual sweep mode; TTL low when dwell is over, or when the point trigger is
received.
NEGative TTL polarity is reversed.
Example
:TRIG:OUTP2:POL NEG
The preceding example sets the signal polarity to be reversed at the TRIG 2 BNC when the trigger is
present.
*RST POS
:TRIGger[:SEQuence]:SLOPe
Supported All
:TRIGger[:SEQuence]:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:SLOPe?
This command sets the polarity of an external signal at the TRIG 1, TRIG 2, or PULSE BNC (see
:TRIGger:EXTernal:SOURce) or internal Pulse Video or Pulse Sync signal (see
:TRIGger:INTernal:SOURce) that will trigger a list or step sweep.
POSitive The signal generator triggers an event when it detects a rising edge on the source
signal
NEGative The signal generator triggers an event when it detects a falling edge on the source
signal
*RST POS
Key Entry Int/Ext Trigger Polarity Neg Pos
Key Entry
Trigger Out 1 Polarity Neg Pos Trigger Out 2 Polarity Neg Pos

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals