108 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Display Subsystem (:DISPlay)
Display Subsystem (:DISPlay)
:ANNotation:AMPLitude[:STATe]
Supported All Models
:DISPlay:ANNotation:AMPLitude[:STATe] ON|OFF|1|0
:DISPlay:ANNotation:AMPLitude[:STATe]?
This command enables or disables the amplitude annotation secure display mode. See also,
“:ANNotation:FREQuency[:STATe]” on page 108 and “:SECurity:DISPlay:RESTricted” on page 185.
On(1) This selection turns off the displayed amplitude security, and the amplitude
annotation is visible.
OFF(0) This selection turns on the displayed amplitude security and the amplitude
annotation is blanked on the signal generator’s display. Also, the keys that access
the amplitude, sweep, and user flatness information are disabled.
For more information about security functions, refer to the User’s Guide.
:ANNotation:AMPLitude:UNIT
Supported All Models
:DISPlay:ANNotation:AMPLitude:UNIT DBM|DBUV|DBUVEMF|V|VEMF|DB
:DISPlay:ANNotation:AMPLitude:UNIT?
This command sets the displayed front- panel amplitude units.
If the amplitude reference state is set to on, the query returns units expressed in dB. Setting any
other unit will cause a setting conflict error stating that the amplitude reference state must be set to
off. Refer to, “:REFerence:STATe” on page 81 for more information.
*RST DBM
:ANNotation:FREQuency[:STATe]
Supported All Models
:DISPlay:ANNotation:FREQuency[:STATe] ON|OFF|1|0
:DISPlay:ANNotation:FREQuency[:STATe]?
This command enables or disables the frequency annotation secure display mode. See also,
“:ANNotation:AMPLitude[:STATe]” on page 108 and “:SECurity:DISPlay:RESTricted” on page 185.
ON (1) This selection turns off the displayed frequency security, and the frequency
annotation is visible.
OFF (0) This selection turns on the displayed frequency security and the frequency
annotation is blanked on the signal generator’s display. Also, the keys that access
the frequency, sweep, and user flatness information are disabled.
For more information about security functions, refer to the User’s Guide.
*RST Activate Restricted Display