EasyManuals Logo

Agilent Technologies N5171B User Manual

Agilent Technologies N5171B
458 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
108 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Display Subsystem (:DISPlay)
Display Subsystem (:DISPlay)
:ANNotation:AMPLitude[:STATe]
Supported All Models
:DISPlay:ANNotation:AMPLitude[:STATe] ON|OFF|1|0
:DISPlay:ANNotation:AMPLitude[:STATe]?
This command enables or disables the amplitude annotation secure display mode. See also,
“:ANNotation:FREQuency[:STATe]” on page 108 and “:SECurity:DISPlay:RESTricted” on page 185.
On(1) This selection turns off the displayed amplitude security, and the amplitude
annotation is visible.
OFF(0) This selection turns on the displayed amplitude security and the amplitude
annotation is blanked on the signal generator’s display. Also, the keys that access
the amplitude, sweep, and user flatness information are disabled.
For more information about security functions, refer to the User’s Guide.
:ANNotation:AMPLitude:UNIT
Supported All Models
:DISPlay:ANNotation:AMPLitude:UNIT DBM|DBUV|DBUVEMF|V|VEMF|DB
:DISPlay:ANNotation:AMPLitude:UNIT?
This command sets the displayed front- panel amplitude units.
If the amplitude reference state is set to on, the query returns units expressed in dB. Setting any
other unit will cause a setting conflict error stating that the amplitude reference state must be set to
off. Refer to, “:REFerence:STATe” on page 81 for more information.
*RST DBM
:ANNotation:FREQuency[:STATe]
Supported All Models
:DISPlay:ANNotation:FREQuency[:STATe] ON|OFF|1|0
:DISPlay:ANNotation:FREQuency[:STATe]?
This command enables or disables the frequency annotation secure display mode. See also,
“:ANNotation:AMPLitude[:STATe]” on page 108 and “:SECurity:DISPlay:RESTricted” on page 185.
ON (1) This selection turns off the displayed frequency security, and the frequency
annotation is visible.
OFF (0) This selection turns on the displayed frequency security and the frequency
annotation is blanked on the signal generator’s display. Also, the keys that access
the frequency, sweep, and user flatness information are disabled.
For more information about security functions, refer to the User’s Guide.
*RST Activate Restricted Display

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals