224 Agilent X-Series Signal Generators SCPI Command Reference
Analog Modulation Commands
Low Frequency Output Subsystem ([:SOURce]:LFOutput)
:SWEep:SHAPe
Supported All Models with Options UNT and 303
[:SOURce]:LFOutput:SWEep:SHAPe SAWTooth|TRIangle
[:SOURce]:LFOutput:SWEep:SHAPe?
This command selects the sweep shape.
SAWTooth A sawtooth sweep shape only sweeps from start freq to stop freq.
TRIangle A triangle sweep sweeps from start freq to stop freq and back to start freq. With
a triangle shape sweep you can choose a different sweep rate for start freq to
stop freq and another sweep rate for stop freq to start freq.
*RST SAWTooth
:SWEep:TIME:COUPled
Supported All Models with Options UNT and 303
[:SOURce]:LFOutput:SWEep:TIME:COUPled ON|OFF|1|0
[:SOURce]:LFOutput:SWEep:TIME:COUPled?
This command sets the couplings between start- to- stop values and stop- to- start values for
low- frequency output sweep times and rates. The sweep times coupled to on.
Use this command when :SWEep:SHAPe is set to TRIangle.
ON|1 This choice uses the same sweep time for both start freq to stop freq and stop
freq to start freq sweeps of a triangle shape sweep. If sweep times coupled is ON,
sweep time and rate is only set by the Start -> Stop softkeys.
OFF|0 This choice disables sweep time couplings.
*RST OFF
:SWEep:TIME[1]|2
Supported All Models with Options UNT and 303
[:SOURce]:LFOutput:SWEep:TIME[1]|2 <val><unit>
[:SOURce]:LFOutput:SWEep:TIME[1]|2?
This command sets the sweep time for the swept- sine waveform.
*RST
+1.00000000Eï€001
Range 1.0mS–65.535S
Key Entry LF Out Sweep Time
Key Entry LF Out Sweep Shape
Key Entry LF Out Sweep Times Coupled Off On