Agilent X-Series Signal Generators SCPI Command Reference 373
Digital Signal Interface Module Commands
Digital Subsystem—Option 003 and 004 ([:SOURce])
This logic type uses single ended I/O and a 3.3 V supply.
Example
:DIG:LOG CMOS15
The preceding example selects 1.5 volt CMOS as the logic data type.
*RST CMOS33
Key Entry
Logic Type
:DIGital:PCONfig
Supported N5172B/82B with option 003 or 004 or both
:DIGital:PCONfig PARallel|SERial|PINTIQ|PINTQI
:DIGital:PCONfig?
This command selects the data transmission type used for communication between the N5102A
module and the device under test.
PARallel This choice selects parallel data transmission.
SERial This choice selects serial data transmission.
PINTIQ This choice selects parallel interleaving data transmission. The I data is
transmitted on the rising clock edge and the Q data on the falling edge.
PINTQI This choice selects parallel interleaving data transmission. The Q data is
transmitted on the rising clock edge and the I data on the falling edge.
Example
:DIG:PCON PINTQI
The preceding example selects parallel interleaving format
*RST PAR
Key Entry
Port Config
:DIGital:PRESet:PTHRough
Supported N5172B/82B with option 003 or 004 or both
:DIGital:PRESet:PTHRough
This command sets up the preset condition for the N5102A module and allows transmission of data
through the module with no modifications. The command is valid only when a modulation format is
active.
Example
:DIG:PRES:PTHR
The preceding example sets the N5102A module to a preset condition and allows data to pass
through unmodified.
Key Entry
Pass Through Preset