Agilent X-Series Signal Generators SCPI Command Reference 49
Basic Function Commands
Frequency Subsystem ([:SOURce])
ON (1) This choice enables the signal generator to detect when a valid reference signal is
present at the 10 MHz IN connector and automatically switches from internal to
external frequency reference.
OFF (0) This choice selects the internal reference oscillator and disables the automatic
switching capability between the internal and an external frequency reference.
*RST 1
Key Entry Ref Oscillator Source Auto Off On
:ROSCillator:OVEN:TUNE
Supported All Models
[:SOURce]:ROSCillator:OVEN:TUNE <value>
[:SOURce]:ROSCillator:OVEN:TUNE?
This command tunes the internal VCTXCO oscillator frequency.
The user value offsets the factory tuned value (the value is added to the factory calibrated DAC
value). The tune value of 0 sets the factory calibrated value.
*RST +0.00000000E+000
Range ï€8192 to +8191
Key Entry Ref Oscillator Tune