Agilent X-Series Signal Generators SCPI Command Reference 151
System Commands
Route Subsystem (:ROUTe)
[:CONNectors]:SOUT
Supported All Models
:ROUTe[:CONNectors]:SOUT SWEep|SETTled|PVIDeo|PSYNc|SW8757|SRUN|SFDone
:ROUTe:CONNectors:SOUT?
This command selects a signal to be routed to the rear panel SWEEP OUT connector.
SWEep This choice routes the sweep out signal to the SWEEP OUT connector.
SETTled This choice routes the source settled signal to the SWEEP OUT connector.
PVIDeo This choice routes the pulse video signal to the SWEEP OUT connector.
PSYNc This choice routes the pulse sync signal to the SWEEP OUT connector.
SW8757 This choice routes the sweep out signal to the SWEEP OUT connector for
compatibility with the 8757D.
SRUN This choice routes the sweep run signal (the pulse at the start of each full sweep)
to the SWEEP OUT connector.
SFDone This choice routes the swept function done signal (the pulse at the end of each
full swept function) to the SWEEP OUT connector.
*RST SWE
Key Entry Route to Sweep Out BNC
[:CONNectors]:TRIGger1:OUTPut
Supported As indicated
:ROUTe[:CONNectors]:TRIGger1:OUTPut
SWEep|SRUN|SETTled|PVIDeo|PSYNc|LXI|PULSe|TRIGger2|SFDone|NONE
:ROUTe[:CONNectors]:TRIGger1:OUTPut?
This command selects a signal to be routed to the rear panel TRIG 1 connector.
SWEep This choice routes the sweep trigger out signal to the TRIG 1 connector.
SETTled This choice routes the source settled signal to the TRIG 1connector.
PVIDeo This choice routes the pulse video signal to the TRIG 1 connector.
PSYNc This choice routes the pulse sync signal to the TRIG 1 connector.
LXI This choice routes the LXI signal to the TRIG 1 connector.
PULSe This choice routes the pulse sync signal to the TRIG 1 connector.
TRIGger2 This choice routes the TRIG 2 BNC signal to the TRIG 1 connector.
SFDone This choice routes the swept function done signal (the pulse at the end of each
full swept function) to the TRIG 1 connector.
NONE This choice routes no signal to the TRIG 1 connector.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals