Agilent X-Series Signal Generators SCPI Command Reference 65
Basic Function Commands
Marker Subsystem–N5173B/83B ([:SOURce])
:MARKer[0]|1|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16|17|18|19:FREQuency
Supported N5173B/83B
[:SOURce]:MARKer[0]|1|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16|17|18|19:FREQuency
<freq>|MAXimum|MINimum
[:SOURce]:MARKer[0]|1|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16|17|18|19:FREQuency?
[MAXimum|MINimum]
This command sets the marker frequency. The frequency value must be between the start and stop
frequencies set for the sweep.
Example
:MARK6 ON
The preceding example turns marker 6 on.
:MARKer[0]|1|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16|17|18|19[:STATe]
Supported N5173B/83B
[:SOURce]:MARKer[0]|1|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16|17|18|19[:STATe]
ON|OFF|1|0
[:SOURce]:MARKer[0]|1|2|3|4|5|6|7|8|9|10|11|12|13|14|15|16|17|18|19[:STATe]?
This command turns a marker on or off. Marker 0 is the default if the marker designator [n] is not
specified.
Example
:MARK6 ON
The preceding example turns marker 6 on.
Range 0 to 19
Key Entry Delta Ref Set
*RST 0
Key Entry Marker On Off
*RST 0
Key Entry Marker On Off