Agilent X-Series Signal Generators SCPI Command Reference 11
SCPI Basics
SCPI Basics
Integer Response Data
Integer response data are decimal representations of integer values including optional signs. Most
status register related queries return integer response data.
The following are examples of integer response data:
Discrete Response Data
Discrete response data are similar to discrete parameters. The main difference is that discrete
response data only returns the short form of a particular mnemonic, in all upper case letters.
The following are examples of discrete response data:
Numeric Boolean Response Data
Boolean response data returns a binary numeric value of one or zero.
String Response Data
String response data are similar to string parameters. The main difference is that string response
data returns double quotes, rather than single quotes. Embedded double quotes may be present in
string response data. Embedded quotes appear as two adjacent double quotes with no characters
between them.
The following are examples of string response data:
0 signs are optional
+100 leading + allowed
100 leading allowed
256 never any decimal point
IMM
EXT
INT
NEG
"This is a string"
"one double quote inside brackets: [""]"
"Hello!"

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals