Model Selection and Configuration
Gem-5 User’s Manual 43
Note behavior of this feature below:
• When dosimeter is removed, the Notify (Clean, Contaminated, Abort, Fault,
and Measurement in progress) signals are all de-activated instead of waiting
for change of screen.
• If Dosimeter is re-inserted on Clean, Contaminated or Go Back screen it re-
triggers prompt to remove dosimeter.
• If Insert Screen is set to "Ready" then removing dosimeter during
measurement re-triggers prompt to insert dosimeter similar to prompt to re-
position hand or foot.
• For LDM61 the Clean Screen displays F3 – Setup, F11 – Instructions,
F4 – Clean – Wait for Read Dosimeter while dosimeter is being read.
• For LDM61 the Contaminated Screen displays F3 – Setup,
F11 – Instructions, F5 – Contaminated – Wait for Read Dosimeter while
dosimeter is being read.
• Note that Notify Signals are active (while dosimeter is inserted) as follows:
− Notify Clean is active while on CLEAN screen.
− Notify Contaminated is active while on CONTAMINATED screen.
− Notify Aborted is active while on GO BACK screen.
− Notify Fault is active while on OUT OF SERVICE screen.
− Measurement in Progress is active while on POSITION, COUNT,
TURN, CLEAN and CONTAMINATED screens.
− Exceptions for LDM61 are
At start of CLEAN the “Notify Measurement in Progress” is left high for
125 ms then set low for 3 s then set high again.
After another 3 s the Remove Dosimeter prompts start but “Notify
Measurement in Progress” remains high for duration of CLEAN screen.
− Exception for LMF is
The Notify Measurement in Progress is set low 125 ms after start of
CLEAN screen and is left low for remainder of sequence.