DMTA-10040-01EN, Rev. E, February 2018
Specifications 337
Conductivity
(
NORTEC 600C,
NORTEC 600S
and
NORTEC 600D)
Frequency 60 kHz or 480 kHz
Digital conductivity
specification
Digital conductivity display from 0.9 % to 110 %
IACS or 0.5 MS/m to 64 MS/m. Accuracy within
± 0.5 % IACS from 0.9 % to 62 % IACS and
within ± 1.0 % of values over 62 %. Meets or
exceeds BAC 5651 specifications. Accuracies
depend on probe frequency, range of calibration,
and coating thickness.
Non-conductive
coating thickness
Nominal accuracy of nonconductive coating
thickness of ± 0.025 mm (± 0.001 in.) from 0 mm
to 0.254 mm (0.00 in. to 0.010 in.) range, and ±
0.50 mm over 10 mm to 0.5 mm (0.01 in. to
0.020 in.) range. Accuracies depend on
conductivity range, probe frequency, and range
of calibration.
Interface Easy to follow conductivity Wizard with
adjustment of reference points and coating
thickness (shims).
Scanners
(NORTEC 600S
and
NORTEC 600D)
Scanner compatibility Operates Olympus scanners (Mini-Mite,
SpitFire, RA-2000, and PS-5) and other major
suppliers’ scanners.
Scanner speed range 40 rpm to 3000 rpm
Display modes Single Impedance, Sweep (strip chart), Split
Screen, and Waterfall.
Waterfall display 60 sweeps per hole
Dual Frequency
(NORTEC 600D)
Frequency adjustment
(dual frequency mode)
Fully independent frequencies, operating in
simultaneous injection
Number of channels 3 (frequency 1, frequency 2, and mix
[NORTEC 600D only])
Display modes New all-in-one display, Dual Impedance, Single
Impedance, Sweep (strip chart), and Split Screen
(sweep + impedance).
MIX options F1 – F2, F1 + F2, and automatic true mixing
Table 8 General and environmental specifications (continued)
Category Parameter Va lue