Wiring
64
FQ2-S/CH User’s Manual
Turning ON the Touch Finder
Power Supply Specifications
Use a power supply that meets the following specifications. (The power supply is sold separately.)
Supply power from a DC power supply for which measures have been applied to prevent high voltages (e.g., a
safety extra-low-voltage circuit).
If UL certification is required for the overall system, use a UL Class II DC power supply.
When using the FQ2-D31, do not connect a switching regulator and AC Adapter (FQ-AC) at the same time.
1 Remove the cover from the power switch on
the left side of the Touch Finder.
2 Press the switch toward ON.
Item Description
Power supply voltage
24 VDC (21.6 to 26.4 V)
Output current
0.65 A min.
Recommended Power Supply
S8VS-01524 (24 VDC, 0.65 A)
External power supply terminal screws
M3.5 (tightening torque: 1.0 Nm)
Recommended power line wire size
AWG16 to AWG22 (length of 5 m max.)