Remote control commandsMXO 4 Series
662User Manual 1335.5337.02 ─ 07
FRANalysis:FREQuency:STARt <StartFrequency>
Sets the start frequency of the sweep.
Parameters:
<StartFrequency> Range: 0.01 to 100000000
Increment: 1
*RST: 100
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Start, Stop" on page 260
FRANalysis:FREQuency:STOP
<StopFrequency>
Sets the stop frequency of the sweep.
Parameters:
<StopFrequency> Range: 0.01 to 100000000
Increment: 1
*RST: 10000
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Start, Stop" on page 260
FRANalysis:GENerator:AMPLitude <GenAmplitude>
Sets a fixed amplitude for the frequency response analysis.
Parameters:
<GenAmplitude> Range: 0.01 to 12
Increment: 0.1
*RST: 1
Default unit: Vpp
Usage: Asynchronous command
Manual operation: See "Amplitude" on page 261
FRANalysis:GENerator:LOAD <GenLoad>
Selects the generator voltage display for 50Ω or high impedance load.
Parameters:
<GenLoad> FIFTy | HIZ
HIZ: high input impedance
*RST: FIFTy
Usage: Asynchronous command
Manual operation: See "User load" on page 261