Remote control commandsMXO 4 Series
663User Manual 1335.5337.02 ─ 07
FRANalysis:GENerator[:CHANnel] <GenCh>
Selects the built-in generator to start a frequency sweep for a defined frequency range.
Parameters:
<GenCh> GEN1 | GEN2
*RST: GEN1
Usage: Asynchronous command
Manual operation: See "Generator" on page 260
FRANalysis:INPut[:SOURce]
<InputChannel>
Sets the channel for the input signal of the DUT.
Parameters:
<InputChannel> C1 | C2 | C3 | C4
*RST: C1
Usage: Asynchronous command
Manual operation: See "Input" on page 260
FRANalysis:MEASurement:DELay:STATe <MeasDelay>
Enables the measurement delay. Set a measurement delay time with FRANalysis:
MEASurement:DELay[:TIME].
Parameters:
<MeasDelay> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Measurement delay, Delay time" on page 264
FRANalysis:MEASurement:DELay[:TIME] <MeasDelayTime>
Sets a time delay, that the system waits before measuring the next point of the plot.
This is helpful in systems that need more time to adapt to the new frequency, for exam-
ple if filters with significant time group delays are present.
Available, if FRANalysis:MEASurement:DELay:STATe = ON.
Parameters:
<MeasDelayTime> Range: 0 to 10
Increment: 0.001
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Measurement delay, Delay time" on page 264