Remote control commandsMXO 4 Series
664User Manual 1335.5337.02 ─ 07
FRANalysis:MEASurement:POINt[:DISPlay] <Points>
Enables the display of the measurement points for the frequency response analysis.
Parameters:
<Points> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Display points" on page 265
FRANalysis:MEASurement:RBW
<RBW>
Sets the resolution bandwidth, which determines the number of measurements that are
used for creating the plot.
Parameters:
<RBW> HIGH | MID | LOW
*RST: MID
Usage: Asynchronous command
Manual operation: See "RBW" on page 264
FRANalysis:OUTPut[:SOURce] <OutputChannel>
Sets the channel for the output signal of the DUT.
Parameters:
<OutputChannel> C1 | C2 | C3 | C4
*RST: C2
Usage: Asynchronous command
Manual operation: See "Output" on page 260
FRANalysis:POINts:LOGarithmic <PtsPerDecade>
Selects the number of points that are measured per decade, if FRANalysis:POINts:
MODE is set to DECade.
Parameters:
<PtsPerDecade> Range: 10 to 500
Increment: 1
*RST: 10
Usage: Asynchronous command
Manual operation: See "Points" on page 261