Remote control commandsMXO 4 Series
683User Manual 1335.5337.02 ─ 07
POWer<m>:QUALity:RESult:POWer:REACtive[:ACTual]?
POWer<m>:QUALity:RESult:POWer:REACtive:AVERage?
POWer<m>:QUALity:RESult:POWer:REACtive:NPEak?
POWer<m>:QUALity:RESult:POWer:REACtive:PPEak?
POWer<m>:QUALity:RESult:POWer:REACtive:RMS?
POWer<m>:QUALity:RESult:POWer:REACtive:STDDev?
Returns the reactive power for the power quality analysis.
For details on the statistics, see "Overview of statistic commands" on page 676.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Return values:
<StdDev> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Usage: Query only
Asynchronous command
POWer<m>:QUALity:RESult:POWer:REALpower[:ACTual]?
POWer<m>:QUALity:RESult:POWer:REALpower:AVERage?
POWer<m>:QUALity:RESult:POWer:REALpower:NPEak?
POWer<m>:QUALity:RESult:POWer:REALpower:PPEak?
POWer<m>:QUALity:RESult:POWer:REALpower:RMS?
POWer<m>:QUALity:RESult:POWer:REALpower:STDDev?
Returns the active power for the power quality analysis.
For details on the statistics, see "Overview of statistic commands" on page 676.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Return values:
<StdDev> Range: -1E+26 to 1E+26
Increment: 1E-10
*RST: 0
Usage: Query only
Asynchronous command
POWer<m>:QUALity:RESult:VOLTage:CREStfactor[:ACTual]?
POWer<m>:QUALity:RESult:VOLTage:CREStfactor:AVERage?
POWer<m>:QUALity:RESult:VOLTage:CREStfactor:NPEak?
POWer<m>:QUALity:RESult:VOLTage:CREStfactor:PPEak?
POWer<m>:QUALity:RESult:VOLTage:CREStfactor:RMS?
POWer<m>:QUALity:RESult:VOLTage:CREStfactor:STDDev?
Returns the voltage crest factor, the Peak value / RMS value for the power quality
analysis.