Remote control commands
R&S
®
RTE
1479User Manual 1326.1032.02 ─ 20
Alternatively, you can set the threshold according to the signal technology with
BUS<m>:UART:TECHnology.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<RxThreshold> User-defined clock threshold
Range: -15 to 15
Increment: 0.1
*RST: 0
Default unit: V
BUS<m>:UART:TX:THReshold <TxThreshold>
Sets a user-defined threshold value for the Tx line.
Alternatively, you can set the threshold according to the signal technology with
BUS<m>:UART:TECHnology.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<TxThreshold> User-defined clock threshold
Range: -15 to 15
Increment: 0.1
*RST: 0
Default unit: V
BUS<m>:UART:TECHnology <Technology>
Sets the threshold voltage Tx and Rx lines as defined for various signal technologies.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Technology> V15 | V25 | V165 | V125 | V09 | VM13 | V38 | V20 | V0 | MANual
V15 | V25 | V165 | V125 | V09 | V38 | V20 | V0
1.5 V, 2.5 V, 1.65 V ... respectively
VM13
-1.3 V (negative value)
MANual
Manual setting of user-defined values with BUS<m>:UART:RX:
THReshold and BUS<m>:UART:TX:THReshold.
*RST: V165