Remote control commands
R&S
®
RTE
1707User Manual 1326.1032.02 ─ 20
TSEQ
Transmission SEQuence: triggers either on the falling edge of
the status nibble, or on the last data nibble.
To set the transmission sequence conditions, use
TRIGger<m>:SENT:TTYPe and TRIGger<m>:SENT:STATus.
To set the data condition, use TRIGger<m>:SENT:TDCN,
BUS<m>:SENT:DNIBbles, TRIGger<m>:SENT:TDMN and
TRIGger<m>:SENT:TDMX.
SMSG
Serial Message: combination of identifier and data conditions.
To select the sequence condition, use TRIGger<m>:SENT:
STYPe.
To select the message ID format for an enhanced serial mes-
sage, use TRIGger<m>:SENT:SIDType.
To set the identifier condition, use TRIGger<m>:SENT:SICN,
TRIGger<m>:SENT:SIMN and TRIGger<m>:SENT:SIMX.
To set the data condition, use TRIGger<m>:SENT:SDCN,
TRIGger<m>:SENT:SDMN and TRIGger<m>:SENT:SDMX.
ERRC
ERRor Condition: triggers on an error event.
Define the error types with TRIGger<m>:SENT:PULSeerror,
TRIGger<m>:SENT:PPERioderror or TRIGger<m>:SENT:
CRCerror.
*RST:
CALI
TRIGger<m>:SENT:TTYPe <TSFieldType>
Selects the trigger sequence type for TRIGger<m>:SENT:TYPE TSEQ (transmission
sequence).
Parameters:
<TSFieldType> STAT | STDA
STAT
Triggers on the status nibble.
STDA
Triggers at the end of the combination of status and data nib-
ble(s).
Define the data conditions with TRIGger<m>:SENT:STATus,
TRIGger<m>:SENT:TDCN, BUS<m>:SENT:DNIBbles,
TRIGger<m>:SENT:TDMN and TRIGger<m>:SENT:TDMX
*RST: STAT
TRIGger<m>:SENT:STATus <StatusBits>
Sets the status nibble data.