Protocol analysis
R&S
®
RTE
530User Manual 1326.1032.02 ─ 20
For "7 bit" and "10 bit", enter the address bits in the Address field, and use the "R/W
bit" on page 530 field to select the transfer direction.
For "7+1 bit", enter the seven address bits and also the R/W bit in the "Address" field.
If the trigger type is "Address + data", you can set the address type "Any" to trigger on
data only, regardless of the address.
Remote command:
TRIGger<m>:I2C:AMODe on page 1430
TRIGger<m>:I2C:ADOR<n>:ADRType on page 1432
Address
The trigger address setup consists of a comparison condition and one or two address
patterns.
Defines the bit pattern of the secondary device address. The length of the entry is
adjusted to the selected address type.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see Chapter 13.1.5, "Bit pattern editor",
on page 521.
"Condition"
Sets the comparison condition to a specific value or a range.
"Min"
Specifies the value or sets the start value of a range.
"Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:I2C:ADDRess on page 1431
TRIGger<m>:I2C:ADDTo on page 1431
TRIGger<m>:I2C:ADOR<n>[:VALue] on page 1432
R/W bit
Toggles the trigger condition between read and write access of the primary. Select
"Either" if the transfer direction is not relevant for the trigger condition.
Remote command:
TRIGger<m>:I2C:ACCess on page 1429
TRIGger<m>:I2C:ADOR<n>:RWBit on page 1432
Monitor
Enables each selected condition, for "Trigger Type" = "Address OR".
Remote command:
TRIGger<m>:I2C:ADOR<n>:ENABle on page 1431
Position
Sets the number of data bytes to be skipped after the address. You can define an exact
position, or a position range. The index 0 is associated with the first data byte.
"Condition"
Sets the comparison condition to a specific value or a range.
"Index min"
Specifies the value or sets the start value of a range.
" Index max"
Sets the maximum value of a range for "Condition" = "In range"/"Out