EasyManuals Logo

Siemens ARCADIS Varic Operator's Manual

Siemens ARCADIS Varic
476 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #234 background imageLoading...
Page #234 background image
Operator Manual
32
of 78 SPR2-310.620.30.02.02
Image Processing
Stopping the movie
â—† Select Scroll > Pause movie or click the button.
– The running movie is stopped.
– If you continue playback of the movie, it will start from the segment where
you paused.
Stopping movie mode
â—† Click the Stop button.
– The running movie is stopped.
– If you continue playback of the movie, it will start from the beginning again.
There is no indication where the film was stopped.
Storing the scene Scenes of a current examination that have not yet been saved can be stored
using the Last Scene Hold function.
Prerequisite: Scene review has been stopped.
â—† Select Patient > Save LSH Scene from the main menu.
—or—
â—† Press function key F9 on the keyboard.
– The last 120 images are stored in the local database.

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors