Operator Manual
26
of 44 SPR2-310.620.30.02.02
System Description
Continuous
Fluoroscopy (CFC)
Press this key.
Fluoroscopy mode is selected.
The LED lights up.
Upon startup of the ARCADIS Varic, the basic setting configured for the body
region "All", which is usually Fluoroscopy, is used as the automatic default
( Register 8: Configuration, Page 5).
Pulsed
Fluoroscopy (PFC)
Press this key.
Pulsed Fluoroscopy mode is selected.
The LED lights up.
Digital
Radiography (DR)
Digital Radiography (DR) provides an electronic instant image with best image
quality. DR is recommended for final exposures.
On activation of digital radiography a short radiation pulse is released.
Press this key.
Digital Radiography mode is selected.
The LED lights up.
In order to ensure sufficient image quality, images are completely acquired after
the start of acquisition, even if no more radiation is released. The effective acqui-
sition time depends on the preset integration factor and is a maximum of 1.8 sec-
onds.
After complete image acquisition, radiation is automatically switched off, even if
the radiation release button remains pressed.