RESULT INTERPRETATION/REPORTS, Stored Data, List Display (P1)
Sysmex SF-3000 Operator's Manual -- Revised September 1995 4-11
NOTE: If the ID bar code reader cannot read a Sample No., an error code
"ERR0000000001" will appear in the Sample No. column, and the "I"
mark will be backlit in red. The error code numbers will go up in
sequence ("...002", "...003", etc.) on subsequent occasions when the
bar code reader cannot read a Sample No.
The code will restart from "ERR0000000001" after the instrument
power turns OFF.
6) RACK
Rack No. of the analyzed sample.
7) TUBE
Mode in which sample was analyzed and sampler analysis tube position No.
A : Sample analyzed in auto (sampler) mode.
C : Sample analyzed in capillary mode.
P : Sample analyzed in closed mode or manual CP mode.
1-10 : Tube position No.
E (backlit in red) : Appears to left of the Tube Position No. Indicates sample
analysis error occurred.
8) OUT
Indicates that the data has not yet been output to a device. Disappears when data is
output.
D (backlit in yellow) : Data Printer
G (backlit in yellow) : Graphic Printer
H (backlit in yellow) : Host Computer
9) POS
Indicates abnormal cell analysis or cell morphology.
There are 4 categories:
D (Diff.) : Indicates abnormality in WBC differential parameters.
M (Morph.) : Indicates abnormal cell morphology.
C (Count) : Indicates abnormality in blood cell count.
Mod (Modified) : Indicates sample of which POSITIVE judgment was changed
manually to NEGATIVE by operator.
10) ERROR
Indicates analysis error. There are two categories:
Func. : Indicates analysis error, excluding the ID read errors and the
sample errors explained below.
Result : Indicates analysis error caused by the sample, such as
"SAMPLE NOT ASP ERROR", "LOW BLOOD VOLUME"
or "LOW COUNT ERROR".