SAMPLE PROCESSING, Manual Mode Procedures
Sysmex SF-3000 Operator's Manual -- Revised September 1995 2-21
(5) When the READY LED turns on again, "Sample Asp Ready" message is
displayed. Prepare the next sample and repeat steps (1)-(4).
NOTE: If sample ID numbers are not set, they are automatically assigned
sequentially.
After the sample is aspirated, the analysis begins. You can check the status of each stage
of analysis using the Analysis Process screen. To see this screen, press [Status] key in
the root menu. Switch to "P2. Analysis Process" using the [→] and [←] keys.
←
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
STATUS
< Analysis Process >
Sequence
ID No.
Rack
Aspirate
Dilute
RBC Count
WBC/BASO Count
DIFF Count
RBC Analyze
WBC/BASO Analyze
DIFF Analyze
Edit
Store
13
123-567-90123
0000-00
12
123-567-90122
0000-00
11
123-567-90121
0000-00
10
123-567-90120
0000-00
: Entry
: Interruption
: Run
: Finish Abnormal
: Finish Normal
→ Return
Figure 2-13: Analysis Process Screen (Manual Mode)
5 . Display Analysis Results
After analysis is completed (approximately 85 seconds after the sample is aspirated), the
results are displayed on the screen.
CAUTION: If all the following conditions are met, the sample is assumed blank and
the analysis result will neither be stored nor judged against any
flagging criteria.
WBC < 1.00 x 10
3
/µL
RBC < 0.30 x 10
6
/µL
HGB < 1.0 g/dL
PLT < 20 x 10
3
/µL