SAMPLE PROCESSING, Auto Mode Procedures
2-30 Sysmex SF-3000 Operator's Manual -- Revised September 1995
CAUTION: • If VENOJECT® II collection tubes are to be used, some engineering
change is required. Contact a Sysmex service representative.
• Recapped VENOJECT® II collection tube cannot be used. Such a
collection tube maybe dropped while mixing or piercing due to a
removed cap.
Use blood sample tubes with a height of 75 mm (overall height including rubber cap is
82 or less) and a diameter of 12, 13, 14 or 15 mm. If tubes have diameters of less than
14 mm, attach adapters to the rack.
Table 2-6: Adapter Requirement
Tube Outer Diameter Adapter
12 mm No. 58
13 mm No. 56
14 mm None
15 mm None
Adapter
Figure 2-20: Adapter
(2) Make sure that the label is in the proper location so that the bar code can be
correctly read. Affix the bar code label as shown in the figures below (tubes and/or
racks).
Also, when setting a sample tube in the rack, make sure that all of the bar codes can
be seen through the rack slits.
Sample Tube
(Sample ID number bar code label)
40mm
16mm
21mm
A
Affix label so that bar
code enters range
indicated by A.
Areas for blood volume
sensor
(Do not affix bar code
label here.)
Rack
(Rack number bar code label)
First Tube
Second Tube
Affix the rack label between the first and the second
tubes, aligning to the top of the rack.
Rack Right
Side
Figure 2-21: Affix Sample Tube Figure 2-22: Affix Rack Bar code
Bar Code Label Label