Commands listed in alphabetical order
Arguments
NR3 is the floati
ng point number that represents the AFG low level value, in volts.
Examples
AFG:LOWLEVEL 1.0 sets the AFG low level value to 1.00 volts.
AFG:LOWLEVEL? might return :AF G:LOWLEVEL -1.5000 indicating the low
level is set to —1.5 Volts.
AFG:NOISEAdd:PERCent
Se
ts (or queries) the AFG additive noise level as a percentage. Minimum is 0.0%,
maximum is 100.0% and increment is 1.0%.
Conditions
Requires option A FG.
Group
AFG
Syntax
AFG:NOISEAdd:PERCent <NR3>
AFG:NOISEAdd:PERCent?
Arguments
Floating point number that represents the AFG additive noise level, as a
percentage.
Examples
AFG:NOISEADD:PERCENT 50 sets the AFG additive noise level to 50 percent.
AFG:NOISEADD:PERCENT? might return :AFG:NOISEADD:PER CENT 0
indicating that no noise is added.
AFG:NOISEAdd:STATE
Sets (or queries) the AFG additive noise state.
Conditi
ons
Requires option A FG.
Group
AFG
Syntax
AFG:NOISEAdd:S
TATE {0|1 |OFF|ON}
AFG:NOISEAdd:STATE?
MSO54, MSO56, MSO58, MSO58LP Programmer 2-103