Commands listed in alphabetical order
BUS:B3:ETHERN
ET:THRESHOLD?
might return
BUS:B3:ETHERNET:THRESHOLD 225.00000E-3 indicates the Bus 3 Ethernet
DATA source High threshold is set to 225.0 mV.
BUS:B<x>:ETHERnet:TYPe
This command species the Ethernet standard speed. The bus number is specied
by <x>.
Conditions
Requires option SR-ENET.
Group
Bus
Syntax
BUS:B<x>:ETHERnet:TYPe {TENBASET|HUNDR EDB
ASETX}
BUS:B<x>:ETHERnet:TYPe?
Arguments
TENBASET species the Ethernet speed as 10Base-T.
HUNDREDBASETX species the Ethernet speed as 100Base-T.
Examples
BUS:B1:ETHERNET:TYPE HUNDREDBASETX species the Ethernet speed as
100Base-T.
BUS:B1:ETHERNET:TYPE? might return TENBASET, indicating that 10Base-T is
the currently specied Ethernet speed.
BUS:B<x>:FLEXray:BITRate
This command sets or queries the FlexRay bus bit rate. The bus is specied by
x. If you select Custom, use BUS:B<x>:FLEXray:BITRate:CUSTom to set the
bit rate.
Conditions
Requires option SR-AUTO.
Group
Bus
Syntax
BUS:B<x>:FLEXray:BITR
ate {CUS TOM|RATE2M|RATE 5M|RATE10M}
BUS:B<x>:FLEXray:BITRate?
Arguments
Arguments specify the bit rate.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-139

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals