Commands listed in alphabetical order
SEARCH:SEARC
H<x>:TRIGger:A:SETHold:CLOCk:THReshold
This command sets or queries the clock threshold setting for a setup/hold trigger
search to determine where to place a mark. The search number is specied by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:SETHold:CLOCk:THReshold <NR3>
SEARCH:SEARCH<x>:TRIGger: A:SETHold:CLOCk:THReshold?
Arguments
<NR3> the clock threshold setting for a setup/hold trigger search.
Examples
SEARCH:SEARCH1:TRIGGER:A: SETHOLD:CLOCK:THRESHOLD - 1.3 sets
the setup/hold trigger clock threshold setting for search 1 to -1.3 volts.
SEARCH:SEARCH1:TRIGGER:A: SETHOLD:CLOCK:THRESHOLD? might return
:SEARCH:SEARCH1:TRIGGER:A :SETHOLD:CLOCK:THRESHOLD –1.3000,
indicating that the setup/hold trigger clock threshold setting for search 1 is set
to -1.3 volts.
** CANN OT USE L ocalCmd HERE BECAUSE OF VARIABLES **
SEARCH:SEARCH<x>:TRIGger:A:SETHold:HOLDTime
This command sets or queries the hold time setting for a s etup/hold trigger search
to determine where to place a mark. The search number is specied by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:SETHold:HOLDTime <NR3>
SEARCH:SEARCH<x>:TRIGger: A:SETHold:HOLDTime?
Arguments
<NR3> species the hold time setting in seconds. Positive values for hold time
occur after the clock edge. Negative values occur before the clock edge.
Examples
SEARCH:SEARCH1:TRIGGER:A: SETHOLD:HOLDTIME 400E-12 sets the
setup/hold trigger hold time setting for search 1 to 400 ps.
SEARCH:SEARCH1:TRIGGER:A: SETHOLD:HOLDTIME? might return
:SEARCH:SEARCH1:TRIGGER:A :SETHOLD:HOLDTIME 500.000 0E–12,
indicating that the setup/hold trigger hold time setting for search 1 is set to 0.5 ns.
2-732 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
Analog Channels4
Sample Rate6.25 GS/s
Vertical Resolution12-bit
Form FactorBenchtop
Probe InterfaceTekVPI
Operating Temperature0 °C to +50 °C
Bandwidth1 GHz
Record Length62.5 Mpoints
Display15.6 inch HD
Waveform Update Rate>500, 000 wfms/s
Trigger TypesEdge, Pulse, Runt, Logic, Serial
Serial Bus SupportI2C, SPI, RS-232/422/485/UART, CAN/CAN FD/LIN, USB 2.0, Ethernet
ConnectivityUSB, Ethernet
Input Impedance1 MΩ
Power Requirements100-240 VAC, 50/60 Hz
Maximum Input Voltage400 V (DC + peak AC)

Related product manuals