Commands listed in alphabetical order
Arguments
<NR3> is the per
centage (where 50% is equal to MID) used to calculate the mid
reference level when the measurement's Ref l evel method is set to Percent.
Examples
MEASUREMENT:CH2:REFLEVELS: PERCENT:FALLMiD 50.0000 sets the MID
reference level for the falling edge to 50%.
MEASUREMENT:CH2:REFLEVELS: PERCENT:FALLMID? might return
:MEASUREMENT:CH2:REFLEVELS :PERCENT:FALLMID 50.0000 indicating the
MID reference level for the falling edge is set to 50%.
MEASUrement:CH<x>:REFLevels:PERCent:HYSTeresis
This command sets or queries the percentage (where 100% is equal to MAX
and 0% is equal to MIN) used to calculate the hysteresis of the reference level
when the measurement ref level method is set to percent. The channel number
is specified by x.
Group
Measurement
Syntax
MEASUrement:CH<x>:REFLevel s:PERCent:HYSTeresis <NR3 >
MEASUrement:CH<x>:REFLevel s:PERCent:HYSTeresis?
Arguments
<NR3> is the hysteresis value used for the autoset.
Examples
MEASUREMENT:CH2:REFLEVELS: PERCENT:HYSTERESIS 5.0000 sets the
reference level hysteresis to 5.0 mV.
MEASUREMENT:CH2:REFLEVELS: PERCENT:HYSTERESIS? might return
:MEASUREMENT:CH2:REFLEVELS :PERCENT:HYSTERESIS 5.000 0 indicating
the reference level hysteresis is set to 5.0 mV.
MEASUrement:CH<x>:REFLevels:PERCent:RISEHigh
This command sets or queries the percentage (where 100% is equal to TOP and
0% is equal to BASE) used to calculate the hig
h reference level of the ris ing edge
when the measurement ref level method is set to percent. The channel number
is specified by x.
Group
Measurement
MSO54, MSO56, MSO58, MSO58LP Programmer 2-389