Commands listed in alphabetical order
Examples
CH3:OFFSET 2.0
E-3
sets the offset for Channel 3 to 2 mV.
CH4:OFFSET? might return :CH4:OFFSET 1.0000E-03, indicating that the
offset for Cha
nnel 4 is set to 1 mV.
CH<x>:POSition
This command sets or queries the vertical position for the specified analog channel.
Group
Vertical
Syntax
CH<x>:POSition <NR1>
Arguments
<NR1> is the vertical position for the specified analog channel.
Examples
CH2:POSition -2.0 sets the position to -2 di
visions.
CH2:POSition? might return :CH2:POSITION -2 .2400 indicating the
position is -2.24 divisions.
CH<x>:PRObe? (Query Only)
This query-only command returns all information concerning the probe that is
attached to the specified channel. The channel is specified by x.
Group
Vertical
Syntax
CH<x>:PRObe?
Related Commands
CH<x>:PROBECal?
Examples
CH2:PROBE? might return :CH2:PROBE:GAIN 1.0000E-01; RESISTANCE
1.0000E+07;UNITS "V";ID:TYPE "10X"'SERNUMBER "N/A"
for a 10X
probe, indicating that (among other parameters) the attenuation factor for the p robe
attached to Channel 2 is 100.0 mV (assuming that probe units are set to volts).
MSO54, MSO56, MSO58, MSO58LP Programmer 2-195